Opto-mechanical assembly and analysis of spectral module in visible-short wave infrared band

ZHANG Quan;LI Xin;ZHAI Wenchao;LIU Enchao;ZHANG Yanna;ZHENG Xiaobing

  • Sponsored by:

    Editor-In-Chief:

    ISSN 1002-2082

     
  • Hosted By:

    Published By: Journal of Applied Optics

    CN 61-1171/O4

Journal of Applied Optics ›› 2019, Vol. 40 ›› Issue (2) : 193-201. DOI: 10.5768/JAO201940.0201003

Opto-mechanical assembly and analysis of spectral module in visible-short wave infrared band

  • ZHANG Quan1,2, LI Xin1, ZHAI Wenchao1, LIU Enchao1, ZHANG Yanna1, ZHENG Xiaobing1
Author information +
History +

Abstract

The spectral modules of self developed radiometer are introduced. The spectral range covers the visible-short wave infrared band(400 nm~2 500 nm).The methods of opto-mechanical assembly of spectral modules are discussed and the results of opto-mechanical assembly are analyzed. The radiometer consists of three spectral modules: the visible spectral module(400 nm~1 000 nm)(VIS for short), the near-infrared spectral module(900 nm~1 700 nm)(NIR for short), and the short wave infrared spectral module(1 600 nm~2 500 nm)(SWIR for short).The detection units of spectral module use the flat-field concave grating to diffract and focus different wavelengths, use the linear array detector to detect signals.The optical units of spectral modules consist of optical fiber, slit, assembly of reflector, assembly of grating and assembly of detector. Since the effect of opto-mechanical assembly is closely related to opto-mechanical design, the result of opto-mechanical assembly can reflect the advantages and disadvantages of opto-mechanical design. According to the characteristics of the opto-mechanical design of the spectral modules, the opto-mechanical assembly was carried out. The results of opto-mechanical assembly show that the wavelength resolution of the three spectral modules is better than 4 nm, 15 nm and 20 nm, respectively. The design indexes of the spectral module are achieved, verifying the rationality of the opto-mechanical design.

Key words

opto-mechanical assembly / visible-short wave infrared / spectral module / opto-mechanical design

Cite this article

Download Citations
ZHANG Quan, LI Xin, ZHAI Wenchao, LIU Enchao, ZHANG Yanna, ZHENG Xiaobing. Opto-mechanical assembly and analysis of spectral module in visible-short wave infrared band. Journal of Applied Optics. 2019, 40(2): 193-201 https://doi.org/10.5768/JAO201940.0201003

References

尹亚鹏, 李新, 郑小兵, 等.场地自动化观测辐射计的设计与实现[J].大气与环境光学学报, 2016, 11(1):44-50.
顾行发, 田国良, 余涛, 等.航天光学遥感器辐射定标原理与方法[M].北京:科学出版社, 2013:199-200.
胡秀清, 张玉香, 邱康睦, 等.中国遥感卫星辐射校正场科研成果论文选编[M].北京:海洋出版社, 2001.
王润生, 熊盛青, 聂洪峰, 等.遥感地质勘查技术与应用研究[J].地质学报, 2011, 85(11): 1699-1743.
林雪松, 王淑荣, 李福田.空间傅里叶变换红外光谱仪用全柔性机构的设计与研究[J].光学精密工程, 2005, 13(6):691-696.
李新, 郑小兵, 洪津, 等.高光谱短波红外地物光谱仪光机设计[J].光学精密工程, 2007, 15(11):1656-1661.
左晓舟, 张燕, 杨海成, 等.量化分析技术在光机装调工艺中的应用[J].新技术新工艺, 2015(1):74-77.
武中华, 袁吕军, 朱永田, 等.特大非球面度离轴非球面补偿器结构设计
CZAPLAMYERS J S, LEISSO N P, ANDERSON N J, et al. On-orbit radiometric calibration of earth-observing sensors using the radiometric calibration test site (RadCaTS)[J]. SPIE, 2012, 8390:83902B-9.
YIN Yapeng, LI Xin, ZHENG Xiaobing, et al. Design and implemeng of automated site observing radiometer[J]. Journal of Atmospheric and Environmental Optics, 2016, 11(1):44-50.
GU Xingfa, TIAN Guoliang, YU Tao, et al.The radiometric calibration principle and method of space remote sensor[M].Beijing:Science Press, 2013:199-200.
SLATER P N, BIGGAR S F, HOLM R G, et al. Reflectance-and radiance-based methods for the in-flight absolute calibration of multispectral sensors[J]. Remote Sensing of Environment, 2015, 22(1):11-37.
HU Xiuqing, ZHANG Yuxiang, QIU Kangmu, et al. Collected papers of China radiometric calibration site[M]. Beijing:Ocean Press, 2001.
WANG Runsheng, XIONG Shengqing, NIE Hongfeng, et al. Remote sensing geology exploration technology and applied research[J]. Acta Geologica Sinica, 2011,

10

Accesses

0

Citation

Detail

Sections