均匀光源被广泛应用于光学成像、遥感仪器的研制与标定等领域。提出了一种具有环形遮挡屏的积分球光源结构,以辐照均匀度分布特性为目标,利用蒙特卡洛(MCM)分析通过改变光源类型、挡板位置、开口大小、光源数量等参数对光源结构进行分析与优化。结果显示,朗伯特光源较之准直型光源辐照均匀度平均提高2.32%;挡板位置靠近积分球边缘时,辐照均匀度略低于其他位置;综合考虑辐射能量与照度均匀度,积分球开口比定在30%~35%区间内更为合适;辐射能量与光源数量呈线性关系,且不均匀度随光源数量在2.46%~3.38%波动。优化后的环形遮挡屏结构,可作为辐照均匀的光源应用于实际场景。
Abstract
The uniform light sources are widely used in optic imaging fields, as well as the development and calibration of remote sensing instruments. For irradiation uniformity, an integrating sphere structure with an annular baffle was proposed. Based on Monte Carlo method, the structure was analyzed and optimized by changing the parameters including light source type, baffle position, slotting size and number of light sources. Results show that the average irradiation uniformity of the Lambert light source is 2.32% higher than the collimated light source. When the baffle is located close to the edge of integrating sphere, the irradiation uniformity is slightly lower than other positions. Considering the radiation energy and illumination uniformity, the open ratio is suggested to be limited in the range of 30%~35%. The radiation power has a linear relationship with the number of light sources, and the nonuniformity fluctuates with the number of light sources from 2.46% to 3.38%. The optimized annular baffle structure is proposed as irradiated uniform light source in practical application scenario.
基金
国家自然科学基金(61604046);贵州省科技计划项目(黔科合平台人才[2017]5788)
{{custom_fund}}
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
刑进, 王淑荣, 李福田.空间紫外遥感光谱辐射计光谱辐亮度定标三种方法的比较[J].中国激光, 2006, 33(4):509-515.
邹思成, 王钢.大天区面积多目标光纤光谱天文望远镜科学级CCD相机低噪声设计和检测[J].兵工学报, 2008, 29(4):434-439.
徐晓峰, 王冀, 张黎明, 等.航天积分球面均匀性检测的研究[J].量子电子学报, 2005, 22(5):817-820.
刘洪兴, 任建伟, 李宪圣, 等.基于LightTools的大口径积分球辐射特性仿真[J].红外与激光工程, 2013, 42(4): 960-965.
张贵彦, 王成, 袁宏韬, 等.积分球出射辐照度的Monte Carlo模拟[J].光电工程, 2006, 33(11): 75-78.
王高明, 郭城, 张亮亮, 等.积分球光能均匀性的Monte Carlo模拟[J].激光与红外, 2009, 39(1):67-69.
颜稳萍, 郭震宁, 林介本, 等.直下式LED平板灯的均匀照明设计[J].光电子·激光, 2014, 25(10):1878-1885.
VEDAT E, SAFAK S, BULENT O. Light sources of solar simulators for photovoltaic devices: a review, renewable and sustainable energy reviews[J]. Renewable and Sustainable Energy Reviews, 2017(77):1240-1250.
XING Jin, WANG Shurong, LI Futian. Comparison of spectral radiance calibrations of spectroradiometer for ultraviolet space remote sensing using three calibration techniques[J]. Chinese Journal of Lasers, 2006, 33(4):509-515.
ZOU Sicheng, WANG Gang. Development and test of scientific CCD camera for large sky area multi-object fiber spectroscopic telescope[J]. Acta Armamentarii, 2008, 29(4):434-439.
XU Xiaofeng, WANG Yi, ZHANG Liming, et al. Space integral spherical uniformity detection[J]. Chinese Journal of Quantum Electronics, 2005, 22(5):817-820.
YANG Chen, Wang Jihong, Guo Xiang, et al. A multisource regular dodecahedron solar simulator structure for uniform Flux[J]. IEEE Journal of Photovoltaics, 2016, 6(2):516-521.
PARK S, LEED H, PARK S N, et al. Experimental validation of the six-p
{{custom_fnGroup.title_cn}}
脚注
{{custom_fn.content}}