摆镜扫描傅里叶变换光谱仪利用摆动扫描方式产生光程差,减小了因动镜倾斜带来的误差影响。从摆扫干涉光谱技术原理出发,对摆镜扫描傅里叶变换光谱仪的光程差进行理论分析,给出了光程差计算表达式,实验结果表明:该光谱仪光程差近似正比于摆动角度。分析了轴外光线因不同入射角所产生的附加光程差,为傅里叶变换光谱仪的指标确定和优化设计提供了理论依据。
Abstract
The optical path difference is an important character in Fourier transform spectrometer, and the optical path difference is produced by means of moving mirrors in spectrometer. Based on the scanning interference spectroscopy principle, the optical path difference of the spectrometer was analyzed theoretically ,and its calculation expression was given .Results show that the optical path difference is approximately proportional to the swing angle.Additional , the optical path differences between beams of light associated with different incident angles were also pointed out. The studies pave the way for optimization design of Fourier transform spectrometer.
关键词
动镜 /
光谱分辨率 /
傅里叶变换 /
光程差
{{custom_keyword}} /
Key words
spectral resolution /
optical path difference /
Fourier transform /
moving mirror
{{custom_keyword}} /
基金
江西省教育厅科学技术研究项目资助(GJJ14316)
{{custom_fund}}
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1]Lu Xuefeng, Bai Qinlan. Optimal design for sagnac interferometer of imaging spectrometer[J]. Journal of Applied Optics,2012, 33(4): 666-669.
逯雪峰,白清兰. 基于成像光谱仪的Sagnac干涉仪的优化设计[J]. 应用光学, 2012, 33(4): 666-669.
[2]Zhou Jinsong, Xiangli Bin, Wei Ruyi, et al. General expression of optic path difference of reflecting rotating Fourier transform spectrometer[J]. Spectroscopy and Spectral Analysis, 2011, 31(1): 263-266.
周锦松,相里斌,魏儒义,等. 反射转镜式干涉光谱仪光程差的一般表达式[J]. 光谱学与光谱分析, 2011, 31(1): 263-266.
[3]Xiangli Bin, Lyu Qunbo, Huang Min. et al. Comparison of two types of Fourier transform imaging spectrometry[J]. Spectroscopy and Spectral Analysis, 2010, 30(5): 1422-1426.
相里斌,吕群波,黄旻,等. 两种干涉成像光谱技术方案比较[J]. 光谱学与光谱分析, 2010, 30(5): 1422-1426.
[4]Yazdi S, Rashidian B. Two new swinging interfer
{{custom_fnGroup.title_cn}}
脚注
{{custom_fn.content}}